Select (Yellow): Select DC/AC in current modes, continuity, diode, capacity, temperature in ohm mode and can also be used to show duty cycle in Hz mode and frequency and duty cycle in AC mode.
Hold: Will freeze the display reading, until pressed again, hold down to activate the background light.
Range, REL: Select manual range and change range, hold down to reactivate automatic ranging. In capacitance it is a REL function to zero out and offset.
Max/min: Capture maximum and minimum values, it will change between max/min/actual, holding it down will activate flashlight.
Rotary switch:
Off: Meter is turned off
VAC: AC voltage and frequency/duty-cycle, the frequency input has high sensitivity and low bandwidth.
VDC: DC voltage
Ohm, diode, continuity, capacitance, temperature Fahrenheit, temperature Celsius
Hz: Logical frequency input, can also show duty cycle.
A: Ampere DC or AC
mA: Milli ampere DC or AC
uA: Micro ampere DC or AC
Test: Phase detection with one connection. This will show 0 or OL in the display, OL means phase detected.
NCV: Non-contact voltage or electric field detection.
10ADC: 10A DC current input.
COM: The common terminal for all ranges.
xxx: All other ranges, including uA and mA.
Measurements
Volt and frequency
1V AC readings is 5% down at 2.3kHz, rms will not work at that frequency
AC frequency range with 0.1V input is from 1.1Hz to 2.5kHz
AC frequency range with 0.1V input it can be stretched to 32kHz
AC frequency range requires a zero crossing.
Frequency input can handle +/-2.8V offset with 1Vrms input.
At 1Vrms frequency input range is from 1.1Hz to 44MHz
Duty cycle works from 1% to 99% at 100kHz with 4Vpp, precision is within 0.3.
Max/min needs about 200ms to capture a voltage, autorange is disabled when using this function and it will not capture every max/min.
Input impedance is 10-11Mohm on DC and AC
Frequency input is 2-3kOhm in input impedance above 2V
Input protection is 1000VDC/750VAC
Frequency input is protected to 250V or 550V DC or AC peak (Both values are specified).
Current
10A range has audible alarm above 10A
10A range is fused with a 10A/250V 5×20mm fuse
mAuA range is fused with a 0.6A/250V 5×20mm fuse
Current ranges starts in DC
Input voltage must not be more than 36VDC or 25VAC
Ohm, continuity, diode and capacitance
Ohm needs about 2.5s to measure 100ohm
Ohm voltage is 1.1V open and 0.25mA shorted.
Continuity is fast (About 20ms).
Continuity beeps when resistance is below 50ohm.
Continuity is 2.1V open and 0.25mA shorted
Diode range uses 4.0V, max. display is 3.3V at 0.25mA, max. current is 1.5mA shorted
10uF takes about 2.5 seconds to measure.
70000uF takes about 7 seconds to measure.
Overload protection is 550V DC or AC peak
Miscellaneous
Current consumption of meter is 1.6mA, 8.2mA with backlight, 9.1mA with backlight and 13.3mA with both.
Meter turns off at 2.2V, battery symbol show at 2.3V.
Readings will be correct until meter turns off.
Backlight fades with dropping voltage, but it usable until meter turns off.
Flashlight fades with dropping voltage, but it usable until meter turns off.
Viewing angle is good except from the top.
Display updates around 3 times/sec
Backlight will automatic turn off after about 12 seconds.
Flashlight will not automatic turn off, but turns off when the meter turns off.
Will automatic turn power off in about 17 minutes.
Standard probes cannot be fully seated in the meter, but they do connect (About 5mm stick up).
The meter usual need a few display update to before it display the value.
Weight is 209g without accessories, but with sleeve and battery.